You're currently on:

GENERAL DESCRIPTION

The SC, XC, & LAB series container is a vacuum insulated container of aluminum with fiberglass neck construction providing you with the highest efficiency possible in liquid nitrogen storage. Use the container for inert fluids only. Liquid oxygen is not compatible with fiberglass material and should not be stored.

A sharp blow to the outer vessel can damage the neck tube or start a vacuum leak. Use caution and common sense in handling the container. Upon receipt of the container, examine it for any evidence of damage during shipping. Watch after the first fill for any signs of vacuum loss, such as frost or sweating on the outside jacket.
(Some frost near the tip just after filling is normal.)

NOTE
Fill the container with a funnel or transfer line when possible. Avoid spilling liquid nitrogen over the vacuum cap near the neck as this can shrink the seal and allow air to leak into the vacuum space.

A. CAUTION (using aluminum SC, XC, LAB series)
To avoid injury by frostbite, use extreme care whenever handling liquid nitrogen, liquid nitrogen storage or transfer vessels or any objects, which have come in contact with liquid nitrogen.
• Leave no areas of skin exposed.
• Always wear proper safety attire over clothing: face shield, cryogenic gloves,
cryogenic apron
• Never overfill liquid nitrogen vessels.
• Always keep liquid nitrogen vessel in an upright position.
• Do not tightly seal liquid nitrogen container or prevent nitrogen gas from escaping.
• Use extreme care to prevent spilling and splashing liquid nitrogen during transfer.
• Immediately remove any clothing or safety attire on which liquid nitrogen has been
spilled.
• Get immediate medical attention for any frostbite injuries due to liquid nitrogen.

B. FILLING INSTRUCTIONS
To avoid damage to your aluminum cryogenic vessel which may result in premature vacuum loss it is important that the following procedure be used during the addition of liquid nitrogen to a warm vessel and on subsequent additions.
1. Slowly pour liquid nitrogen to new or warm vessels.
2. Allow liquid nitrogen to sit in covered vessel for 2 hours to completely cool inner
3. Fill your vessel to the desired level after the 2 settling (cooldown) time.
4. If you are filling your dewar from a pressurized source, make sure that the source
tank is at a low pressure (22 PSI or below).
5. If transfer hose is used for extracting liquid nitrogen from a pressurized liquid source
always use a phase separator on the end of the hose.
6. Remember to always wear proper safety attire over clothing; face shields cryogenic
gloves and apron.
7. Never overfill your dewar with liquid nitrogen. Overfilling the tank may cause
immediate or premature vacuum failure to occur.

C. MEASURING LIQUID NITROGEN QUANTITY
1. Use wooden or plastic dipstick. Never use a hollow tube to measure liquid nitrogen.
2. Level will be indicated by frostline, which develops when dipstick is removed.

D. LIQUID WITHDRAWAL
1. liquid withdrawal for the LAB units is always done by pouring or utilizing a
withdrawal device. Withdrawal device pressurizes to approximately 5 psi and the
pressure forces liquid up the withdrawal tube out the valve.
2. Always wear proper safety attire; shield, gloves and apron.

REPLACEMENT PARTS
MODELS SC3/3 SC8/5 SC11/7 SC16/11 SC20 SC36
Canister 9710601 9710611 9710091 9721489 9710101 9710101
Cork/cover 10507059 10507059 10507059 10507438 10726817 1072681
Pumpout caps 3911217 3911217 3911217 3911217 3911217 3911217
MODELS XC20/20 XC21/6 XC22/5 XC32/8
XC33/22 XC34/18
Canister 11006344 9721469 9719349 9719339 9719319 9719309
Cork/cover 11028236 10507024 10506996 10507454 10507067 10507489
Pumpout caps 3911217 3911217 3911217 3911217 3911217 3911217
MODELS XC35/12 XC43/28 XC47/11-6 XC47/11-10 XC47/11-6 SQ.
Canister 10854966 9719319 9719299 9719289 9723199
Cork/cover 10855723 10507067 10721397 10726711 10721397
Pumpout caps 3911217 3911217 3911217 3911217 3911217
MODELS LAB 4 LAB 5 LAB 10 LAB 20 LAB
30 LAB 50
Cork./cover 10588362 10580299 10580299 10580475 10580459 10580459
Pumpout caps 3911217 3911217 3911217 3911217 3911217 3911217

WARNING: the venting of nitrogen vapors will create a dilution of the air’s oxygen concentration necessary to support life. Exposure to this diluted atmosphere can cause asphyxiation or even death. DO NOT store or use liquid container in areas that have poor ventilation. Place liquid container outdoors or in a well-ventilated area. Failure to comply with this warning may cause serious personal injury including death.